Sofia Escobar Londono
Comparative evaluation of PPG-based heart-rate estimation across three wearable devices.
Rel. Luigi Borzi', Gabriella Olmo, Irene Rechichi. Politecnico di Torino, Corso di laurea magistrale in Ingegneria Biomedica, 2026
|
Preview |
PDF (Tesi_di_laurea)
- Tesi
Licenza: Creative Commons Attribution Non-commercial No Derivatives. Download (18MB) | Preview |
Abstract
Wearable photoplethysmography (PPG) is widely used for heart-rate estimation outside clinical settings, where body movement and posture are not controlled. This thesis investigates the performance and agreement of three wearable PPG devices under conditions representative of activities of daily living. The study compares Persimmon, a research prototype worn on the forearm, with a Polar Verity Sense worn on the proximal forearm and a Shimmer3 GSR+ positioned at the earlobe, with the aim of characterizing their heart-rate estimates across different levels of physical activity and postural change. Twenty-two laboratory recordings were conducted, during which each participant simultaneously wore all three devices. The experimental protocol comprised seated rest, three seated tasks, unsupported standing rest, walking, and postural transitions.
Optical signals from all three devices were processed using the same heart-rate estimation pipeline, thereby minimizing differences attributable to device-specific heart-rate estimation algorithms
Relatori
Anno Accademico
Tipo di pubblicazione
Numero di pagine
Corso di laurea
Classe di laurea
URI
![]() |
Modifica (riservato agli operatori) |