Martina Retico
A 200 kHz Inductive Wireless Power Receiver for Active Implantable Medical Devices.
Rel. Paolo Motto Ros, Andrea Dentis. Politecnico di Torino, Corso di laurea magistrale in Ingegneria Biomedica, 2026
|
Preview |
PDF (Tesi_di_laurea)
- Tesi
Licenza: Creative Commons Attribution Non-commercial No Derivatives. Download (13MB) | Preview |
Abstract
Wireless Power Transfer (WPT) is emerging as a promising technology for powering implantable medical devices (IMDs), offering advantages in terms of absence of wired connections, miniaturization, safety and system integration. This thesis focuses on the design and development of a high-efficiency receiver (RX) circuit for a Near-Field Inductive Coupling (NIRC) WPT system, operating at a low resonant frequency of 200 kHz. The WPT system is intended to wirelessly power an implanted bidirectional peripheral nervous system interface (RHS2116), designed for the control of an upper limb mechatronic exoprosthesis. The overall architecture of the WPT system consists of a transmitter (TX) circuit integrated within the prosthetic socket and the receiver implanted in the patient’s arm.
The main design challenges include the need to meet miniaturization requirements imposed by implantations constrains and to maintain efficient power transfer under coil misalignment conditions, which can occur due to patient movement and prothesis shifting