Luana Puleo
Smart Glasses system for the maculopathy evaluation based on Pattern-Reversal Visual Evoked Potentials.
Rel. Danilo Demarchi, Alessandro Sanginario, Rossana Terracciano. Politecnico di Torino, Corso di laurea magistrale in Ingegneria Biomedica, 2019
|
Preview |
PDF (Tesi_di_laurea)
- Tesi
Licenza: Creative Commons Attribution Non-commercial No Derivatives. Download (28MB) | Preview |
Abstract
• Objective: the aim of this work is to test the performance of the bio-signal amplifier g.HIamp Research in the ophthalmological field, by comparison with the current commercial device used by Retimax CSO. In detail, the study is aimed at the collection and analysis of pr-VEPs, reversal patterns Visual Evoked Potentials, bio-potentials obtained through appropriate visual stimulation of the subject. • Materials & Methods: the two devices use two different stimulation systems: in particular, at g.HIamp Research, the Epson Moverio BT-200 Smart glasses, a portable and economical visual stimulation system, were added. The EEG signals, obtained by stimulation and taken by electrodes positioned according to standard 10/20, are sent to the bio-amplifier that records and processes them through the software g.Recorder associated with it.
This is followed by further Matlab processing that highlights the characteristics of the waves obtained, in terms of amplitude and latency
Relatori
Tipo di pubblicazione
URI
![]() |
Modifica (riservato agli operatori) |
